Output Characteristics
Modes | Maximum P-P Voltage | Rated Load (OHMS) | Maximum Power (Watts) | Crest Factor* (Typical) | |
Monopolar Cut | Pure | 900 | 500 | 400 | 1.5 |
Blend I | 1100 | 500 | 300 | 2.1 | |
Blend II | 1200 | 500 | 250 | 2.4 | |
Blend III | 1300 | 500 | 200 | 2.8 | |
Blend IV | 1300 | 500 | 150 | 3.1 | |
Monopolar Coagulation | Soft | 1600 | 500 | 250 | 3.5 |
Desiccate | 1800 | 500 | 200 | 3.8 | |
Force | 2100 | 500 | 150 | 4.2 | |
Spray | 3000 | 500 | 100 | 6.8 | |
Bipolar | Micro | 320 | 100 | 90 | 1.5 |
Standard | 460 | 100 | 90 | 1.5 | |
Marco / Cut | 600 | 100 | 90 | 1.5 |
Diathermy GX 400
Description
PPQM System
- Measuring Frequency : 85kHz ± 10kHz
- Measuring Current : <10μA
- Dual Area PREM Patient Return Electrode : 5-135Ω
- Single Area Patient Return Electrode : <10Ω
Features
- PEQM : Patient Electrode Contact Quality Monitoring Safety against the Patient Return Electrode site burns.
- Programmable : User can Program as per Surgical Procedure requirements.
- SPRAY Coagulation : Homogeneous, superficial, faster Coagulation over larger area.
- Bipolar Cut : Marco mode can be used for bipolar cutting with hooks and other cutting devices.
- Desiccate Coagulation : Faster coagulation for thicker tissues.
- Foot Switches : Explosion Proof and Water Proof Foot Switch.
- Time Out Error : Automatically stops the HF output as Time Limit Exceeds.
Applications Fields
Ophthalmic, Laparoscopic, Surgery, Endo, GI, Ortho, Open & General Surgery, Oncology, Gynecology, Hernia, GI Endocytosis, Papillotomy, Polypectomy, Under Water, Urology, Gynecology TCRE, Hysteroscopy, Thoracotomy, Unit is suitable for all kind of Open, Endoscopic, Laparoscopic and Minimally Invasive Surgery Plastic, Neuro, ENT, Spine, Thoracic and Cardiac Surgery.
Safety Standards
- Classification According to the EC – Directive 93 / 42 / EEC. Class llb.
- General Safety Standards : IEC 60601-1-1, IEC 60601-2-2
- EMI/EMC Compatibility Standard : IEC 60601-1-2
- IS Safety Standard : IS 7583
Leakage Current
- HF Leakage : Monopolar <150mA RMS-Bipolar <60Ma RMS.
- Low Frequency : Normal Polarity, intact ground : <10µA.
- Normal Polarity, ground open : <50µA
Input Power
- Operating Voltage : 190-260Volt 50Hz AC, (Optional 90-120Volts 60Hz AC)
Dimensions
- L x W x H – 350 x 350 x 130mm
- Net Weight in 5 Kg.