Diathermy Touch FX-400x
Output Characteristics
Modes | Maximum P-P Voltage | Rated Load (OHMS) | Maximum Power (Watts) | Crest Factor* (Typical) | |
Monopolar Cut | Pure | 900 | 500 | 400 | 1.5 |
Blend I | 1100 | 500 | 300 | 2.1 | |
Blend II | 1200 | 500 | 250 | 2.4 | |
Blend III | 1300 | 500 | 200 | 2.8 | |
Blend IV | 1300 | 500 | 150 | 3.1 | |
Monopolar Coagulation | Soft | 1600 | 500 | 300 | 3.5 |
Desiccate | 1800 | 500 | 250 | 3.8 | |
Force | 2100 | 500 | 200 | 4.2 | |
Spray | 3000 | 500 | 120 | 6.8 | |
Bipolar | Micro | 320 | 100 | 70 | 1.5 |
Standard | 460 | 100 | 90 | 1.5 | |
Marco / Cut | 600 | 100 | 120 | 1.5 |
DESCRIPTION
VESSEL SEALING MODE : Pure Sine wave 440 kHz Seal – Pulse output mode,
Soft-Seal 120 W at 100 CF1.5
Med-Seal 200 W at 100 CF1.5
Rapid-Seal 350 W a t 100 CF1.5
PPQM System
Measuring Frequency : 85 kHz ± 10 kHz
Measuring current : <10 μA
Dual Area Patient Return Electrode : 5-135 Ω
Application Fields
Opthalmic, Laparo Surgery, GI, Ortho, Open & General Surgery, Onco, Gynecology, Hernia, GI Endocut, Papillotomy, Polypectomy, Under Water, Urology, Gynec TCRE, Hysteroscopy, Thoracotomy, Unit is suitable for all kind of Open, Endoscopic, Laparoscopic and Minimally Invasive Surgery Plastic, Neuro, ENT, Spine, Thoracic and Cardiac Surgery.
Safety Standards
Classification according to the EC-Directive 93/ 42/ EEC. Class llb General Safety Standards : IEC 60601-1-1, IEC 60601-2-2 EMI/EMC Compatibility Standard : IEC 60601-1-2 IS Safety Standard : IS 7583.
Leakage Current
HF Leakage Monopolar <100 mA RMS-Bipolar <50 mA RMS
Low frequency : Normal polarity, intact ground : < 10 μA
Normal polarity, ground open : < 50 μA
INPUT POWER
Operating Voltage : 190-260 Volt 50Hz AC, (Optional 90-120Volts 60Hz AC)
DIMENSIONS : L x W x H – 350 x 350 x 130mm Net Weight in 605 Kg.