Diathermy Touch SX-400xc
Output Characteristics
Modes | Maximum P-P Voltage | Rated Load (OHMS) | Maximum Power (Watts) | Crest Factor* (Typical) | |
Monopolar Cut | Pure | 900 | 500 | 400 | 1.5 |
Blend I | 1100 | 500 | 300 | 2.1 | |
Blend II | 1200 | 500 | 250 | 2.4 | |
Blend III | 1300 | 500 | 200 | 2.8 | |
Blend IV | 1300 | 500 | 150 | 3.1 | |
Monopolar Coagulation | Soft | 1600 | 500 | 250 | 3.5 |
Desiccate | 1800 | 500 | 200 | 3.8 | |
Force | 2100 | 500 | 150 | 4.2 | |
Spray | 3000 | 500 | 100 | 6.8 | |
Bipolar | Micro | 320 | 100 | 70 | 1.5 |
Standard | 460 | 100 | 90 | 1.5 | |
Marco / Cut | 600 | 100 | 120 | 1.5 |
Description
SALINE RESECTION (BITUR) MODE: Sine wave 440 kHz
Saline Cut. – Saline Plasma Cutting Variable 10 W to 300 W at 100 Ù, CF 1.5
Saline Coag- Saline Plasma Coag Variable 10 W to 150 W at 100 Ù, CF 1.5
PPQM System
Measuring Frequency : 85 kHz ± 10 kHz
Measuring current : < 10 μA
Dual Area PREM Patient Return Electrode : 5-135 Ù
Single Area Patient Return Electrode : < 10 Ù
Applications Fields
Opthalmic, Laparo Surgery, Endo, GI, Ortho, Open & General Surgery, Onco, Gynecology, Hernia, GI Endocut, Papillotomy, Polypectomy, Under Water, Urology, Gynec TCRE, Hysteroscopy, Thoracotomy, Unit is suitable for all kind of Open, Endoscopic, Laparoscopic and Minimally Invasive Surgery Plastic, Neuro, ENT, Spine, Thoracic and Cardiac Surgery.
Safety Standards
Classification according to the EC- Directive 93/ 42/ EEC: Class IIb
General Safety Standards: IEC 60601-1-1, IEC 60601-2-2
EMI/EMC Compatibility Standard: IEC 60601-1-2
IS Safety Standard: IS 7583
Leakage Current
HF Leakage: Monopolar <150 mA RMS- Bipolar <60 mA RMS
Low frequency : Normal polarity, intact ground : < 10 μA
Normal polarity, ground open : < 50 μA
INPUT POWER
Operating Voltage : 190-260 Volt 50Hz AC, (Optional 90-120Volts 60Hz AC)
DIMENSIONS : L x W x H – 350 x 350 x 130mm Net Weight in 6.5Kg