Infant Radient Warmer

The Infant Radient Warmer is mounted on wheels for easy mobility with fixed baby cradle and totally epoxy coated finish.

Technical Specifications

> Heater Box Consists of Ceramic Infra Red Heater 650 Watts.
> Due to temprature insulated cover on the heater box, its exterior surface remains cool even after long run.
> The construction of the stainless steel reflactor and guard rail is such that the complete cleaning of heater & reflector is possible without any electrical shock hazards Halogen Examination Lamp (12 Volts, 50 Watts) / High Intensity LED lamp to provide accurate assessment of the infant color and sufficient illumination during procedures Baby Tray.
> Clear acrylic collapsible side panels.
> Baby Tray with transparent acrylic sheet to incorporate facility of under surface phototheraphy.
> Acrlic Tray and cushion mattress.
> The design aspects also allow easy of dismantling baby care area during cleaning or fumigation.

  Option 1 Option 2
Control Modes Skin / Air / Manual Skin / Manual
Range Skin Control : 34 to 38° C
Air Control : Room Temprature to 39° C
Skin Control : 34 to 38° C
Resolution 0.1° C
Accuracy ± 0.2° C
Display Bright Skin & Air Temp.(7 Segment LED) Bright Skin (7 Segment LED)
% of heater output Bar Graph
Mannual Mode Heater output controllable from 0 to 100% in increments of 5% in manual mode
Timer 0 to 60 mins 0 to 60 mins
Maual Mode 1. Time Programmable
2. Heater output settable
3. Over Temp. cut off
Safety Alarm with Mute Facility 1. Skin Mode
    Skin High: 1.0° C
    Skin Low: 1.0° C
    Under Range: <34° C
    Over Range >38° C
2. Air Mode :
    Air High : 1.0° C
    Over Range
3. Heater Failure
4. Power Control Failure
5. Probe & Power Failure Alarm
1. Skin High: 1.0° C
2. Skin Low: 1.0° C
3. Under Range: <34° C
4. Over Range >38°
5. Heater Failure
6. Power Control Failure
7. Probe & Power Failure Alarm

Available with choice of :

> Lower Shelf and 3 Drawer Storage Cabinet.
> Side Mounted Monitor Tray.
> Pole for Syringe Pump / Infusion Pump.
> Integrated baby bassinet with trendlenburg & Reverse Facility.
> The Source box can be swiveled to take X-ray.
> Acrylic X-ray Tray with X-ray Guide for positioning the Cassette.
> LCD Display : Control Mode & Set temp, Time & Alarm message.
> APGAR Timer.
> Integrated / Suctioning / Oxygen flow & regulation Facility.
> Oxygen Cylinder & Cylinder Attachments.
> Height Adjustable Facility
> Battery Back-up Facility
> Integrated Photo Theraphy Unit.